EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Preset Commands

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
356User Manual 1178.3711.02 ─ 08
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage: Event
12.4 Preset commands
The preset commands are not bundled in one subsystem. Therefore, they are listed
separately in this section.
Four presetting actions are available:
Activating the default state of all internal instrument functions (*RST on page 354).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. reference oscillator source settings.
Activating the preset state of the parameters related to the selected signal path (:
SOURce<hw>:PRESet on page 356)
Activating the preset state of all parameters that are not related to the signal path
(:DEVice:PRESet on page 356)
Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 357). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
When resetting, the following deviation between remote and manual control exists:
In contrast to the [Preset] key, the SCPI commands *RST and :SYSTem:PRESet do
not close open dialogs in the GUI.
:DEVice:PRESet............................................................................................................ 356
:SOURce<hw>:PRESet.................................................................................................. 356
:SYSTem:PRESet.......................................................................................................... 357
:SYSTem:FPReset......................................................................................................... 357
:DEVice:PRESet
Presets all parameters which are not related to the signal path, including the LF gener-
ator.
Example:
DEV:PRES
Presets all instruments settings that are not related to the signal
path.
Usage: Event
:SOURce<hw>:PRESet
Presets all parameters which are related to the selected signal path.
Preset commands

Table of Contents

Other manuals for Rohde & Schwarz R&S SMB100B

Related product manuals