Analog modulations
R&S
®
SMB100B
102User Manual 1178.3711.02 ─ 08
"Trapezoid"
1 = Trapezoid rise
2 = Trapezoid high
3 = Trapezoid period
4 = Trapezoid fall
Remote command: 
[:SOURce<hw>]:LFOutput<ch>:SHAPe on page 468
Frequency
Sets the frequency of the LF generator for sine signals.
Set the signal shape with the parameter Shape.
Remote command: 
[:SOURce]:LFOutput<ch>:FREQuency on page 464
Period
Sets the repetition rate of the generated LF signal for triangle, trapezoid or pulse
shapes, see Shape.
The period of sine signals is calculated from the selected Frequency
Remote command: 
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:PERiod on page 470
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:PERiod on page 470
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:PERiod on page 469
Pulse Width
Sets the pulse duration of the generated pulse signal.
Remote command: 
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:WIDTh on page 469
Pulse Duty Cycle
Sets the ratio between the pulse duration and the pulse period in percent.
Remote command: 
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:DCYCle on page 468
Modulation settings