EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Page 469

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
469User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:PERiod <Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.
Parameters:
<Period> float
Range: 1E-6 to 100
Increment: 1E-8
*RST: 1E-3
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Period" on page 102
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:WIDTh
<Width>
Sets the pulse width of the generated pulse.
Parameters:
<Width> float
Range: 1E-6 to 100
Increment: 1E-8
*RST: 5E-4
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Pulse Width" on page 102
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:FALL <Fall>
Selects the fall time for the trapezoid shape of the LF generator.
Parameters:
<Fall> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Trapezoid Rise / Fall" on page 103
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:HIGH <High>
Sets the high time for the trapezoid signal of the LF generator.
Parameters:
<High> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
SOURce subsystem