EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Page 470

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
470User Manual 1178.3711.02 ─ 08
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Trapezoid High" on page 103
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:PERiod <Period>
Sets the period of the generated trapezoid shape. The period determines the repetition
frequency of the internal signal.
Parameters:
<Period> float
Range: 1E-6 to 100
Increment: 1E-8
*RST: 1E-3
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Period" on page 102
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:RISE
<Rise>
Selects the rise time for the trapezoid shape of the LF generator.
Parameters:
<Rise> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Trapezoid Rise / Fall" on page 103
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:PERiod <Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.
Parameters:
<Period> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.001
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Period" on page 102
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:RISE <Rise>
Selects the rise time for the triangle single of the LF generator.