Remote control commands
R&S
®
SMB100B
465User Manual 1178.3711.02 ─ 08
Usage: Query only
[:SOURce<hw>]:LFOutput:FREQuency:MANual <Manual>
Sets the frequency of the subsequent sweep step if LFO:SWE:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any value within the setting range, where:
STARt is set with [:SOURce<hw>]:LFOutput:FREQuency:
STARt
STOP is set with [:SOURce<hw>]:LFOutput:FREQuency:
STOP
Range: STARt to STOP
Increment: 0.1
*RST: 1000
Example: See Example"Setup an LF sweep" on page 461.
Manual operation: See "Current Frequency" on page 130
[:SOURce<hw>]:LFOutput:FREQuency:MODE <Mode>
Sets the mode for the output of the LF generator frequency, and determines the com-
mands to be used for frequency settings.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Sets the fixed-frequency mode. CW and FIXed are synonyms.
To set the output frequency, use command [:SOURce]:
LFOutput<ch>:FREQuency
SWEep
Sets sweep mode.
To set the frequency, use the commands:
[:SOURce<hw>]:LFOutput:FREQuency:STARt and [:
SOURce<hw>]:LFOutput:FREQuency:STOP
Or [:SOURce<hw>]:LFOutput:FREQuency:MANual
*RST:
CW
Example: See Example"Setup an LF sweep" on page 461.
Manual operation: See "State (LF frequency sweep)" on page 100
SOURce subsystem