Remote control commands
R&S
®
SMB100B
466User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:LFOutput:FREQuency:STOP <Stop>
[:SOURce<hw>]:LFOutput:FREQuency:STARt <Start>
Sets the start/stop frequency for [:SOURce<hw>]:LFOutput:FREQuency:MODE
SWEep.
Parameters:
<Start> float
Range: 0.1 Hz to 1 MHz
Increment: 0.1
*RST: 1 KHz
Example: See Example"Setup an LF sweep" on page 461.
Manual operation: See "Start Frequency/Stop Frequency" on page 135
[:SOURce]:LFOutput<ch>[:STATe]
<State>
Activates LF signal output.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "State" on page 106
[:SOURce]:LFOutput<ch>:INTernal:VOLTage <Voltage>
Sets the output voltage for the LF generators.
The sum of both values must not exceed the overall output voltage, set with command
[:SOURce]:LFOutput:VOLTage.
Suffix:
<ch>
.
[1]|2
LF1 and LF2
Parameters:
<Voltage> float
Range: 0 to 4
Increment: 0.001
*RST: 1
Example: See Example"Configuring the LF generator" on page 462.
Manual operation: See "Output Voltage" on page 106
[:SOURce]:LFOutput:OFFSet <Offset>
Sets a DC offset at the LF Output.
SOURce subsystem