15.3.1.1 Internal adjustment settings
Access:
► Select "System Config > Setup > General > Internal Adjustments".
In this dialog, you can perform internal calibration routines, and get information on
the last performed calibration.
The "Temperature Offset" indicates the deviation of the current temperature of the
instrument, compared to the temperature of the last adjustment.
The remote commands required to define these settings are described in Chapter 12.6,
"CALibration subsystem", on page 367.
How to: See Chapter 15.3.1.3, "How to use the internal adjustments", on page 590.
Settings
Adjust All
Performs all available internal calibration routines of the instrument.
How to: See Chapter 15.3.1.3, "How to use the internal adjustments", on page 590.
Remote command:
:CALibration:ALL[:MEASure]? on page 368
Last Full Adjustment
Displays the date of the last fully performed adjustment.
Remote command:
:CALibration<hw>:ALL:DATE? on page 368
Time
Displays the elapsed time since the last full adjustment.
Performing maintenance tasks