Remote control commands
R&S
®
SMB100B
368User Manual 1178.3711.02 ─ 08
:CALibration<hw>:ALL:TIME?......................................................................................... 369
:CALibration:DATA:FACTory:DATE?.................................................................................369
:CALibration<hw>:CONTinueonerror................................................................................370
:CALibration:DELay:MINutes...........................................................................................370
:CALibration:DELay:SHUTdown[:STATe].......................................................................... 370
:CALibration:DELay[:MEASure]?..................................................................................... 370
:CALibration:ALL[:MEASure]? [<Force>]
Starts all internal adjustments that do not need external measuring equipment.
Query parameters:
<Force> string
Return values:
<Measure> 1 | ON | 0 | OFF
Example:
CAL:ALL:MEAS?
// Response: "0"
// Executes the adjustments of all instrument functions.
// When completed, it indicates that the adjustment
// has been performed successfully.
Usage: Query only
Manual operation: See "Adjust All" on page 588
:CALibration<hw>:ALL:DATE?
Queries the date of the most recently executed full adjustment.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Date> string
Example:
CAL:ALL:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Full Adjustment" on page 588
:CALibration<hw>:ALL:INFormation?
Queries the current state of the internal adjustment.