Remote control commands
R&S
®
SMB100B
480User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:LIST:INDex:STARt <Start>
[:SOURce<hw>]:LIST:INDex:STOP <Stop>
Sets the start and stop index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list.
Parameters:
<Stop> integer
Index range
Only values inside this range are processed in list mode
Range: 0 to list length
*RST: 0
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Manual operation: See "List Range from/to" on page 142
[:SOURce<hw>]:LIST:MODE <Mode>
Sets the list mode.
The instrument processes the list according to the selected mode and trigger source.
See LIST:TRIG:SOUR AUTO, SING or EXT for the description of the trigger
source settings.
Parameters:
<Mode> AUTO | STEP | INDex
AUTO
Each trigger event triggers a complete list cycle.
STEP
Each trigger event triggers only one step in the list processing
cycle. The list is processed in ascending order.
INDex
The trigger event triggers the entry of the selected index in the
list.
*RST: AUTO
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Manual operation: See "Mode" on page 140
[:SOURce<hw>]:LIST:POWer <Power>
Enters the level values in the selected list. The number of level values must corre-
spond to the number of frequency values. Existing data is overwritten.
Parameters:
<Power> <Power#1>{, <Power#2>, ...} | block data
SOURce subsystem