Remote control commands
R&S
®
SMB100B
455User Manual 1178.3711.02 ─ 08
CW|FIXed
Sets the fixed frequency mode. CW and FIXed are synonyms.
The instrument operates at a defined frequency, set with com-
mand [:SOURce<hw>]:FREQuency[:CW|FIXed].
SWEep
Sets sweep mode.
The instrument processes frequency (and level) settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:FREQuency:STARt on page 458 and [:
SOURce<hw>]:FREQuency:STOP on page 459,
[:SOURce<hw>]:FREQuency:CENTer on page 458,
[:SOURce<hw>]:FREQuency:SPAN on page 458,
[:SOURce<hw>]:FREQuency:MANual on page 457
LIST
Sets list mode.
The instrument processes frequency and level settings by
means of values loaded from a list.
To configure list mode settings, use the commands of the Chap-
ter 12.15.6, "SOURce:LIST subsystem", on page 474.
COMBined
Sets the combined RF frequency / level sweep mode.
The instrument processes frequency and level settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:COMBined:FREQuency:STARt
on page 513 and [:SOURce<hw>]:COMBined:FREQuency:
STOP on page 514,
[:SOURce<hw>]:COMBined:POWer:STARt on page 514 and
[:SOURce<hw>]:COMBined:POWer:STOP on page 514
*RST:
CW
Example: See Example"Frequency configuration" on page 454,
and Example"Setup an RF frequency or power sweep"
on page 506
Manual operation: See "State (RF frequency sweep)" on page 129
See "State (RF frequency / level sweep)" on page 130
See "State" on page 140
[:SOURce<hw>]:FREQuency[:CW|FIXed] <Fixed>
Sets the frequency of the RF output signal in the selected path.
The effect depends on the selected mode:
●
In CW mode (FREQ:MODE CW | FIXed), the instrument operates at a fixed fre-
quency.
●
In sweep mode (FREQ:MODE SWE), the value applies to the sweep frequency.
The instrument processes the frequency settings in defined sweep steps.
SOURce subsystem