Remote control commands
R&S
®
SMB100B
458User Manual 1178.3711.02 ─ 08
See "RF frequency and level display with a downstream instrument" on page 59.
Note: The offset also affects RF frequency sweep.
Parameters:
<Offset> float
Increment: 0.01
*RST: 0
Example: See Example"Frequency configuration" on page 454
Manual operation: See "Offset" on page 61
[:SOURce<hw>]:FREQuency:CENTer <Center>
Sets the center frequency of the sweep.
See Chapter 6.2.1, "Correlating parameters in sweep mode", on page 122.
Parameters:
<Center> float
Range: 300 kHz to RFmax
Increment: 0.01 Hz
*RST: 300E6
Default unit: Hz
Example: See Example"Setup an RF frequency or power sweep"
on page 506
Manual operation: See "Center Frequency" on page 135
[:SOURce<hw>]:FREQuency:SPAN <Span>
Sets the sapn of the frequency sweep range.
See Chapter 6.2.1, "Correlating parameters in sweep mode", on page 122.
Parameters:
<Span> float
Full freqeuncy range
Increment: 0.01
*RST: 400E6
Example: See Example"Setup an RF frequency or power sweep"
on page 506
Manual operation: See "Span" on page 135
[:SOURce<hw>]:FREQuency:STARt <Start>
Sets the start frequency for the RF sweep.
See Chapter 6.2.1, "Correlating parameters in sweep mode", on page 122.
SOURce subsystem