Remote control commands
R&S
®
SMB100B
450User Manual 1178.3711.02 ─ 08
Example: See Example"Fill user correction data with sensor" on page 447.
Usage: Event
Manual operation: See "Fill User Correction Data with Sensor" on page 165
[:SOURce<hw>]:CORRection:CSET[:SELect] <Filename>
Selects or creates a file for the user correction data.
If the file with the selected name does not exist, a new file is created.
Parameters:
<Filename> string
Filename or complete file path; file extension can be omitted.
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Manual operation: See "UCOR Data" on page 160
See "Functions for handling of data lists" on page 208
[:SOURce<hw>]:CORRection:VALue?
Queries the current value for user correction.
Return values:
<Value> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Usage: Query only
Manual operation: See "User Correction" on page 160
[:SOURce<hw>]:CORRection:ZERoing:STATe <State>
Activates the zeroing procedure before filling the user correction data acquired by a
sensor.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 1
Example: See Example"Fill user correction data with sensor" on page 447.
Manual operation: See "Fill User Correction Data with Sensor" on page 165
SOURce subsystem