Remote control commands
R&S
®
SMB100B
451User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Fills the selected user correction list with the level values measured by the power sen-
sor for the given frequencies.
Suffix:
SENSor<ch>
.
Defines the used power sensor, i.e. the sensor whose values are
used.
Example: See Example"Fill user correction data with sensor" on page 447.
Usage: Event
[:SOURce<hw>]:CORRection[:STATe]
<State>
Activates user correction with the currently selected table.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Manual operation: See "State" on page 160
[:SOURce]:CORRection:CSET:CATalog?
Queries a list of available user correction tables.
Return values:
<Catalog> string
List of list filenames, separated by commas
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Usage: Query only
Manual operation: See "UCOR Data" on page 160
See "Directory, File List and Filename" on page 203
[:SOURce]:CORRection:CSET:DELete <Filename>
Deletes the specified user correction list file.
Setting parameters:
<Filename> string
Filename or complete file path; file extension is optional.
Example: See Example"Create a table with user-defined correction values
for the RF level" on page 446 .
Usage: Setting only
SOURce subsystem