Remote control commands
R&S
®
SMB100B
510User Manual 1178.3711.02 ─ 08
Usage: Query only
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] <Logarithmic>
Sets a logarithmically determined step size for the RF level sweep. The level is
increased by a logarithmically calculated fraction of the current level.
See Chapter 6.2.1, "Correlating parameters in sweep mode", on page 122.
Parameters:
<Logarithmic> float
The unit dB is mandatory.
Range: 0.01 to 139 dB
Increment: 0.01
*RST: 1
Default unit: dB
Example: See Example"Setup an RF frequency or power sweep"
on page 506.
Manual operation: See "Step" on page 137
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl <Dwell>
Sets the dwell time for a frequency sweep step.
Parameters:
<Dwell> float
Range: 0.001 to 100
Increment: 100E-6
*RST: 0.01
Example: See Example"Setup an RF frequency or power sweep"
on page 506.
Manual operation: See "Dwell Time" on page 133
[:SOURce<hw>]:SWEep[:FREQuency]:MODE <Mode>
Sets the cycle mode for the frequency sweep.
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Each trigger event triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually by input of the frequencies with the command [:
SOURce<hw>]:FREQuency:MANual.
SOURce subsystem