EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Page 510

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
510User Manual 1178.3711.02 ─ 08
Usage: Query only
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] <Logarithmic>
Sets a logarithmically determined step size for the RF level sweep. The level is
increased by a logarithmically calculated fraction of the current level.
See Chapter 6.2.1, "Correlating parameters in sweep mode", on page 122.
Parameters:
<Logarithmic> float
The unit dB is mandatory.
Range: 0.01 to 139 dB
Increment: 0.01
*RST: 1
Default unit: dB
Example: See Example"Setup an RF frequency or power sweep"
on page 506.
Manual operation: See "Step" on page 137
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl <Dwell>
Sets the dwell time for a frequency sweep step.
Parameters:
<Dwell> float
Range: 0.001 to 100
Increment: 100E-6
*RST: 0.01
Example: See Example"Setup an RF frequency or power sweep"
on page 506.
Manual operation: See "Dwell Time" on page 133
[:SOURce<hw>]:SWEep[:FREQuency]:MODE <Mode>
Sets the cycle mode for the frequency sweep.
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Each trigger event triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually by input of the frequencies with the command [:
SOURce<hw>]:FREQuency:MANual.
SOURce subsystem

Table of Contents

Other manuals for Rohde & Schwarz R&S SMB100B

Related product manuals