Remote control commands
R&S
®
SMB100B
550User Manual 1178.3711.02 ─ 08
EAUTo
An external signal triggers the sweep. As soon as one sweep is
finished, the next sweep starts. A second trigger event stops the
sweep at the current frequency and level value pairs, a third trig-
ger event starts the trigger at the start values, and so on.
*RST: No enum range assigned!
Manual operation: See "Mode" on page 130
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>:LFFSweep:IMMediate
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>[:SWEep][:IMMediate]
Performs a single sweep and immediately starts the activated, corresponding sweep:
FSWeep - RF frequency
PSWeep - RF level
LFFSweep - LF frequency
SWEep - all sweeps
Effective in the following configuration:
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR SING
SOUR:SWE:FREQ|POW:MODE AUTO or SOUR:LFO:SWE:[FREQ:]MODE AUTO
Alternatively, you can use the IMMediate command instead of the respective SWEep:
[FREQ:]|POW:EXECute command.
Example:
TRIG
Starts all active sweeps.
Usage: Event
Manual operation: See "Execute Single Sweep" on page 134
:TRIGger<hw>:LFFSweep
Executes an LF frequency sweep in the following configuration:
TRIG:LFFS:SOUR SING
LFO:SWE:MODE AUTO
Example:
LFO:SWE:MODE AUTO
TRIG:LFFS:SOUR SING
TRIG:LFFSweep
Usage: Event
TRIGger subsystem