Remote Control Commands
R&S
®
SMB100A
311Operating Manual 1407.0806.32 ─ 13
LFOutput:FREQuency:STOP 10 kHz
2. Select linear or logarithmic sweep spacing.
LFOutput:SWEep[:FREQuency]:SPACing LIN
3. Set the step width and dwell time.
LFOutput:SWEep[:FREQuency]:STEP[:LINear] 100 Hz
LFOutput:SWEep[:FREQuency]:DWELl 20 ms
4. Determine the sweep mode.
LFOutput:SWEep:MODE AUTO
5. Determine the trigger.
TRIGger0:SOURce SINGle
6. Activate the sweep.
LFOutput:FREQuency:MODE SWEep
7. Trigger the sweep (depending on the mode).
LFOutput:SWEep:EXECute
[:SOURce]:LFOutput<ch>:FREQuency.............................................................................311
[:SOURce]:LFOutput:FREQuency:MANual.......................................................................312
[:SOURce]:LFOutput:FREQuency:MODE.........................................................................312
[:SOURce]:LFOutput:FREQuency:STARt.........................................................................313
[:SOURce]:LFOutput:FREQuency:STOP..........................................................................314
[:SOURce]:LFOutput[:STATe]..........................................................................................314
[:SOURce]:LFOutput:SWEep[:FREQuency]:DWELl...........................................................314
[:SOURce]:LFOutput:SWEep[:FREQuency]:EXECute........................................................314
[:SOURce]:LFOutput:SWEep[:FREQuency]:MODE............................................................315
[:SOURce]:LFOutput:SWEep[:FREQuency]:POINts...........................................................316
[:SOURce]:LFOutput:SWEep[:FREQuency]:RUNNing?......................................................316
[:SOURce]:LFOutput:SWEep[:FREQuency]:SHAPe...........................................................316
[:SOURce]:LFOutput:SWEep[:FREQuency]:SPACing........................................................317
[:SOURce]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]................................................317
[:SOURce]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic..........................................318
[:SOURce]:LFOutput:SHAPe...........................................................................................319
[:SOURce]:LFOutput:SIMPedance...................................................................................319
[:SOURce]:LFOutput:VOLTage........................................................................................319
[:SOURce]:LFOutput<ch>:FREQuency <Frequency>
The command sets the frequency of the LF signal for CW mode (:SOUR:MODE CW). The
setting is valid for all analog modulations (AM/FM/PhiM) with internal modulation source
and for the LF Output.
In sweep mode (SOUR:LFO:FREQ:MODE SWE), the frequency is linked to the sweep
frequency.
Correlation:LFO:FREQ for LFO:FREQ:MODE SWE linked to sweep frequency.