Remote Control Commands
R&S
®
SMB100A
312Operating Manual 1407.0806.32 ─ 13
Parameters:
<Frequency> float
Increment: 0.1 Hz
*RST: 1000
Example:
LFO2:FREQ 5kHz
sets the frequency of the LF generator 2 signal to 5 kHz.
Manual operation: See "LF Gen Frequency" on page 171
[:
SOURce]:LFOutput:FREQuency:MANual <Manual>
In Sweep mode (SOUR:LFO:FREQ:MODE SWE) the command sets the frequency for the
next sweep step in the "Step" sweep mode (SOUR:LFO:SWE:MODE MAN). Here only
frequency values between the settings SOUR:LFO:FREQ:STAR and ...:STOP are per-
mitted. Each sweep step is triggered by a separate SOUR:LFO:FREQ:MAN command.
Parameters:
<Manual> float
Range: STARt to STOP
Increment: 0.1 Hz
*RST: 1 kHz
Example:
LFO:SWE:MODE MAN
sets the "Step" sweep mode.
LFO:FREQ:MAN 5 kHz
sets an LF frequency of 5 kHz for the next step in the "Step" sweep
mode.
LFO:FREQ:MODE SWE
sets the LF Sweep mode. An LF frequency of 5 kHz is output.
LFO:FREQ:MAN 5.1 kHz
triggers the next sweep step with a frequency of 5.1 kHz.
Manual operation: See "Current Freq - LF Sweep" on page 192
[:SOURce]:LFOutput:FREQuency:MODE <Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output frequency.
SOURce Subsystem