Remote Control Commands
R&S
®
SMB100A
429Operating Manual 1407.0806.32 ─ 13
Example:
STEReo:DIRect? "RDS-PH"
Response: "100"
STEReo:DIRect "RDS-DEV=<Deviation>"
STEReo:DIRect? "RDS-DEV"
(for documentation reasons only)
Sets the RDS frequency deviation (max. deviation).
Use the SCPI command [:SOURce]:STEReo:RDS[:DEViation] instead.
Setting parameters:
<Deviation> Note: A four-digit value must always be set. Leading zeros, if any,
must also be specified.
Range: 0000 to 1000 (ASCII coded decimal numbers), cor-
responding to 0 Hz to 10.00 kHz)
Example:
STEReo:DIRect "RDS-DEV=0201"
The RDS frequency deviation is set to 2.01 kHz.
Example:
STEReo:DIRect? "RDS-DEV"
Response: "0201"
STEReo:DIRect "RDS-PRESET"
All RDS specific parameters are deleted or set to a default values.
Example:
STEReo:DIRect "RDS-PRESET"
Sets all RDS parameter to their preset values
Usage: Event
STEReo:DIRect "RT=<RetranNumber>,<A/
BFlag>,<RadioTextMsg#1>,<RadioTextMsg#2>"
STEReo:DIRect? "RT"
Radio text
Setting parameters:
<RetranNumber> Range: 00 to 15 (ASCII coded decimal numbers), number
of retransmissions of radio text message
<A/BFlag> 0 | 1
If the A/B flag is set, the A/B bit in group 2A is toggled to signal
that a new radio text message will be transmitted.)
Direct Commands for the Stereo/RDS Coder Option R&S SMB-B5