EasyManua.ls Logo

STONEX R2 - 3.3 Start Survey

STONEX R2
220 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STONEX R2
29
The calculating formula for atmospheric correction is as follows: (calculating
unitmeter)
PPM = 273.8 - 0.2900 ³ Pressure ValuehPa
1 + 0.00366 × Temperature value(‵
If the pressure unit adopted is mmHg: make conversion with:
1hPa = 0.75mmHg.
The standard atmospheric condition of R SERIES Total Station instrument (e.g. the
atmospheric condition under which the atmospheric correction value of the instrument is
zero):
Pressure1013 hPa
Temperature: 20
If regardless of atmospheric correction, please set PPM value as 0.
OPERATIONAL STEPS
OPEROPERATION OPERATION
DISPLAY
In the screen of EDM setting, press
[F1] (Atmos) to enter atmospheric
correction function.
[F1]
EDM Settings
EDM Mode Tracking
Reflect: Prism
Prism 30.0mm
ATMOS GRID SET
The current settings display on the
screen, use navigation key to move
the cursor to Temp item.
Atmosphere Data
Retr.Corr 0.14
Temp 20
Pressure 1013.2 hPa
Atmos PPM 0 PPM
BACK PPM=0 SET
Input temperature value,
e.g.: Input 26, press [ENT] to move
the cursor to Pressure item.
Input 26
+
[ENT]
Atmosphere Data
Retr.Corr 0.14ATUR
Temp 26
Pressure 1013.2 hPa
Atmos PPM 0 PPM
BACK PPM=0 SET

Table of Contents

Related product manuals