4
Life Science Applications
Pierce Modified Lowry
208 Evolution 200 Series User Guide Thermo Scientific
Related Topics
Type Tab for Pierce Modified Lowr y
Measurement Tab for Pierce Modified Lowry
Standards Tab for Pierce Modified Lowry
Instrument Tab for Pierce Modified Lowry
Accessories Tab for Pierce Modified Lowry
Lowry Standard Curves
A standard curve is required for colorimetric protein analysis.
• A standard curve can be generated using two or more standards.
• The multipoint curve capability allows multiple replicates for each standards. Standards
can be run in any order; however, best practice dictates that the standards be measured
from the lowest concentration to the highest.
• Determine whether or not to use a standard when analyzing samples against the standard
curve.
• Establishing a new standard curve requires creating a new workbook.
• If a previously saved workbook is used, all concentration calculations for newly measured
samples will be based on the standard absorbance values saved in the workbook.
Related Topics
Overview
Application Settings
Lowry Standard Curves
Making Pierce Modified Lowry Assay Measurements
Making Pierce Modified Lowry Assay Measurements
See Sample and Standards Preparation for more information.
Note Each workbook will archive only one standard curve.