Peltier Thermostatted Single Cell Holder
316 Evolution 200 Series User Guide Thermo Scientific
Description of Symbols
Environmental Conditions
Performance and safety of the accessory are guaranteed if operated under the following
conditions:
• Pollution degree 2 (standard CEI EN 61010-1).
• Installation category II with smaller transient over voltages (standard IEC 664).
• Mains voltage fluctuations must not exceed 10% of the nominal voltage.
• Location of use is indoors at an altitude <2000 m (6562 ft).
• Location of use exhibits a relative humidity <80% noncondensing.
• Location of use exhibits an ambient temperature between 10 °C and 35 °C (50 °F and
95 °F).
• Location of use is free from dust and vapors of solvents and acids.
Related Topics
Temperature Controller Front and Rear Panels
General Safety Information
Preinstallation
Installation
Operation
Maintenance
Troubleshooting
Preinstallation
Review this section before installing the accessory.
CAUTION Symbol identifies the use of substances dangerous for health and the
environment. For these substances, refer to the laws and terms in force for the laboratory
using the instrument. Take the necessary precautions when using flammable or explosive
substances.