Mercury Lamp
Thermo Scientific Evolution 200 Series User Guide 273
Mercury Lamp
This mercury lamp can be used to calibrate and verify the performance of your UV-Visible
spectrophotometer.
Safety Considerations
The mercury lamp has a safety interlock to prevent accidental exposure to dangerous levels of
UV radiation during operation. After the lamp is installed, removing the lamp from the port
while the instrument is powered on will activate the interlock to block all radiation from the
lamp.
To avoid defeating the safety interlock, keep all magnets and strong magnetic fields (such as
those produced by a mass spectrometer or various medical instruments) away from the lamp
while the instrument is powered on and the lamp cable is connected.
Consignes de sécurité
La lampe à vapeur de mercure comprend un verrouillage de sécurité destiné à prévenir toute
exposition accidentelle à des doses dangereuses de rayonnement ultraviolet pendant le
fonctionnement de l’instrument. Une fois la lampe installée, et lorsque l’instrument est sous
tension, le retrait de la lampe du port active le verrouillage afin de bloquer le rayonnement de
la lampe.
Pour ne pas déjouer le verrouillage de sécurité, tenez les aimants et autres champs magnétiques
puissants (tels que ceux produits par un spectromètre de masse ou divers instruments
médicaux) éloignés de la lampe lorsque l’instrument est sous tension et que le câble de la
lampe est raccordé.