25
Smart Linear 8-Cell Changer
Thermo Scientific Evolution 200 Series User Guide 365
Y To clean the accessory exterior
1. Remove the accessory from the sample compartment. See Installing a Smart Accessory.
2. Use a cloth dampened with plain water to wipe the surface of the Smart Accessory base
and cell holder.
3. Dry the surface with another cloth.
Related Topics
Routine Maintenance
Maintenance
Troubleshooting
Troubleshooting
Related Topics
Safety Considerations
Operating Precautions
Installation
Operation
Maintenance
Problem Possible Cause Solution
Absorbance
values are
inconsistent or
higher than
expected
The accessory is not
aligned properly
Calibrate and then initialize the cell changer.
See Calibrating the Cell Changer and
Initializing the Cell Changer.