4
Life Science Applications
Pierce 660 nm Protein Assay
214 Evolution 200 Series User Guide Thermo Scientific
Related Topics
Measurement Tab for Pierce 660 nm Protein Assay
Standards Tab for Pierce 660 nm Protein Assay
Instrument Tab for Pierce 660 nm Protein Assay
Accessories Tab for Pierce 660 nm Protein Assay
Samples Tab for Pierce 660 nm Protein Assay
Measurement Tab for Pierce 660 nm Protein Assay
Feature Description
Name Application name.
Description (optional) Description of the template.
Standard curve Type of quantitative analysis used.
Pathlength Distance the light travels through the sample.
Click Settings in Pierce 660 nm Protein Assay to display
the Measurement tab in the right pane.
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
To use uncorrected data, select None. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant, the
calculated protein concentration may be higher than the true value.
Single point subtracts the sample absorbance at the entered
wavelength from the absorbance at every data point.
Sloping baseline draws a line between the data points at the entered
wavelengths, finds its absorbance value at the analytical wavelength,
and subtracts it from the measured absorbance.
Component name Component to quantify.