Protein Biuret
230 Evolution 200 Series User Guide Thermo Scientific
Related Topics
Type Tab for Protein Biuret
Measurement Tab for Protein Biuret
Standards Tab for Protein Biuret
Instrument Tab for Protein Biuret
Accessories Tab for Pierce 660 nm Protein Assay
Standard Curves
A standard curve is required for colorimetric protein analysis.
• A standard curve can be generated using two or more standards.
• The multipoint curve capability allows multiple replicates for each standards. Standards
can be run in any order; however, best practice dictates that the standards be measured
from the lowest concentration to the highest.
• Determine whether or not to use a standard when analyzing samples against the standard
curve.
• Establishing a new standard curve requires creating a new workbook.
• If a previously saved workbook is used, all concentration calculations for newly measured
samples will be based on the standard absorbance values saved in the workbook.
Related Topics
Overview
Application Settings
Making Protein Biuret Assay Measurements
Making Protein Biuret Assay Measurements
See Sample and Standards Preparation for more information.
Y To make Protein Biuret assay measurements
1. Click Measure Protein Biuret in Protein Biuret.
Note Each workbook will archive only one standard curve.