29
Temperature Probe
434 Evolution 200 Series User Guide Thermo Scientific
Related Topics
Operating Precautions
Installation
Maintenance
Troubleshooting
Temperature Settings
In the selected application, use Settings > Accessories > Temperature to display the available
settings for a temperature probe or controller. The Temperature features appear only when the
driver for a temperature probe or controller is installed.
Related Topics
Configuring a Target Temperature
Displaying a Temperature Status Monitor
Temperature Settings
Configuring a Target Temperature
Displaying a Temperature Status Monitor
Feature Description
Record temperature with mea-
surement
If a temperature probe is used, records the temperature of
the sample after each measurement. Temperature readings
are stored in the sample properties.
Use temperature controller If a Peltier cell holder or changer is used, defines a setpoint
temperature for the cell block and a stir speed if desired.
The available settings depend on the specifications for the
temperature-controlled accessory.
Start measurements at target
temperature
Starts the measurement when the sample (requires a tem-
perature probe) or cell block (requires a Peltier cell holder
or changer) reaches a specified temperature.
Match probes to cell changer
positions
If a temperature probe is used, maps probe channels 1
through 8 with up to eight cell changer positions. Useful
only when using an automatic cell changer with multiple
temperature probes.
Deselect this to always record the temperature of probe 1.