EasyManua.ls Logo

Thermo Scientific Evolution 201 - CVC Test Limits; Running Performance Verification Tests; Using the CVC

Thermo Scientific Evolution 201
500 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Calibrated Validation Carousel (CVC)
Thermo Scientific Evolution 200 Series User Guide 253
CVC Test Limits
The pass/fail limits are based on the instrument specification and any other tolerances
associated with the Certified Reference Material (CRM). The software stores all tolerances
and any instrument pass/fail limits with the calibration data and reports these values as
appropriate.
Related Topics
Installing a Smart Accessory
Installing the Carousel on the Smart Base Unit
Installing Complete CVC in the Instrument
Loading CVC Calibration Data
Using the CVC
Maintenance
Using the CVC
Before using the CVC, mount the CVC carousel on the base unit and install the accessory in
the spectrophotometer. See Installing the Carousel on the Smart Base Unit and Installing
Complete CVC in the Instrument.
If using this CVC for the first time, load the calibration data. See Loading CVC Calibration
Data.
Running Performance Verification Tests
The CVC performance tests appear only when the CVC is properly installed and a valid
calibration file has been loaded.
Y To run CVC performance verification tests
1. Click Performance Verification in Home.
The available tests appear in the right pane.
Contents
Running Performance Verification Tests
Viewing CVC Test Reports
Note If you are running security software, you can perform performance verification tests
only if you have been given the right to do so by the system administrator.

Table of Contents

Related product manuals