12
Calibrated Validation Carousel (CVC)
252 Evolution 200 Series User Guide Thermo Scientific
CVC Test Limits
Using the CVC
Maintenance
Loading CVC Calibration Data
Each CVC has unique calibration data shipped with the accessory. Before a CVC can be used
for performance testing, it’s serial number and calibration file must be read by INSIGHT
software. It is not necessary to reload the data unless you purchase a new CVC or have it
recalibrated.
Make a backup copy of the calibration data before use. Store the master and the Certificate of
Calibration in a secure location.
Y To load CVC calibration data
1. From the INSIGHT software Home screen, click Performance Verification.
2. Click Load CVC Calibration. A dialog box lets you browse for the calibration file. The
file is provided on a memory stick or other medium. The following format is used for the
file name:
evodata_[x]_[y].cal
where: x is the serial number of the CVC carousel, and
y is the CVC type (e for EP, u for USP and s for Standard).
3. Locate and select the calibration file that matches the installed CVC and click OK.
Related Topics
Installing a Smart Accessory
Installing the Carousel on the Smart Base Unit
Initialization
CVC Test Limits
Using the CVC
Maintenance
Note The software can store one calibration file for each type of CVC.