EasyManua.ls Logo

Thermo Scientific Evolution 201 - Cleaning the Accessory; Maintenance; Routine Maintenance

Thermo Scientific Evolution 201
500 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
SPE-8W Smart Peltier Linear Cell Changer
392 Evolution 200 Series User Guide Thermo Scientific
12. Place the cuvettes in the correct cell changer positions.
13. Click OK or press Run on the instrument keypad.
To cancel, click Cancel or press Stop on the instrument keypad.
If a blank is required, it will be measured first, followed by the samples in order.
If the sample list has more entries than there are spaces in the cell changer, the Loading
Guide reappears after the last cuvette is measured.
14. If the loading guide appears, remove all cuvettes from the cell changer and fill it with the
next set of samples as defined in the loading guide.
15. Click OK.
16. Continue as above until all entries in the sample list have been measured.
Related Topics
Calibrating the Cell Changer
Initializing the Cell Changer
Setting the Target Temperature and Stir Speed
Configuring the Cell Changer
Operation
Maintenance
Note If the application requires a blank and it was not measured in step 4, the first
cell changer position will be measured as the blank.
Contents
Routine Maintenance
Cleaning the Accessory

Table of Contents

Related product manuals