Loading...
20
Peltier Thermostatted Single Cell Holder
Thermo Scientific Evolution 200 Series User Guide 315
• Stirring speed
Displayed Values
• Peltier block temperature
• Set temperature
• Stirring speed
• Heating/cooling power
Operating Specifications
• Stirring speed 0 to 1800 rpm
• Temperature range: 0°C to 100°C (32°F to 212°F)
• Temperature accuracy: ±0.15°C (±0.27°F) from 0°C to 100°C (32°F to 212°F)
• Temperature repeatability: typically ±0.1 °C (±0.18 °F)
• Temperature stability: typically ±0.1 °C (±0.18 °F)
• Specifications valid for ambient temperatures of 15 °C to 35 °C (59 °F to 95 °F)
External Control
•RS-232
•USB
Dimensions
• Width: 213 mm (8.3 in)
• Height: 178 mm (7 in)
• Depth: 269 mm (10.6 in)
• Weight: 3 kg (8 lbs)
Electrical Data
• Voltage: 115/120 V or 220/240 V ±10%
• Frequency: 50/60 Hz
• Power: 80 VA
•CE marked