INSIGHT Software
Operation
Thermo Scientific Evolution 200 Series User Guide 67
Samples Tab for Quant
Related Topics
Quant
Type Tab for Quant
Measurement Tab for Quant
Standards Tab for Quant
Instrument Tab for Quant
Accessories Tab for Quant
Formulas and Units Tab
Quantifying Sample Data
Performing a Quantitative Analysis
Click Settings in Quant to display the Samples tab in
the right pane.
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Tr ip li ca te , respectively.
Use concentration
limits
Displays high and low limit lines on the Run Chart tab to show
whether sample concentrations are within the specified limits.
Use weight/volume
correction
If available for the current unit, corrects sample concentrations using
the entered target weight and volume:
corrected concentration = measured concentration * (actual weight /
target weight) * (target volume / actual volume)