2 MEASurement Commands
82 U2000 Series Programming Guide
1
The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the
command sub-systems. The parameters must be entered in the specified order. If parameters are
omitted, they default from the right. The parameter DEFault is used as a place holder. Specifying
DEF leaves the parameter value unchanged.
2
When the measurement result is linear this parameter represents the number of significant
digits. When the measurement result is logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001
respectively.
Example
source list The measurement which the command is
implemented on.
If unspecified the current window setup is
used.
(@1)
Item Description/Default Range of Values
MEAS1:POW:AC?
-70DBM,1,(@1)
This command queries the current
window/measurement of the sensor, using
an expected power level of - 70 dBm and a
resolution setting of 1.