8SERVice Subsystem
209 U2000 Series Programming Guide
SERVice:SENSor[1]:CPLace <“place”>
This query is used to enter the calibration place.
Syntax
Example
Query
SERVice:SENSor[1]:CPLace?
This query returns the calibration place in U2000 Series USB power sensors.
Calibration place information is stored in the sensor’s non- volatile memory.
Query example
SERV:SENS1:CPL
“Agil-Penang”
This command enters the place of
calibration of the U2000 Series USB power
sensor as Agilent Penang.
SERV
1
:CPL
:SENS
“place”
?
Space
SERV:SENS1:CPL? This query returns the place of
calibration of the U2000 Series USB Power
Sensor.