8SERVice Subsystem
211 U2000 Series Programming Guide
SERVice:SENSor[1]:FREQuency:MAXimum?
This query returns the maximum frequency that can be measured in MHz unit
by the currently connected sensor. It is applicable to U2000 Series USB power
sensors only. Maximum frequency information is stored in the sensor’s
non- volatile memory.
Syntax
Example
Error Messages
If the sensor, currently connected, does not contain the necessary
information in non- volatile memory, error –241 “Hardware missing”
occurs.
SERV:SENS1:FREQ:MAX? This query returns the maximum
frequency that can be measured by the
U2000 Series USB Power Sensor
currently.
SERV
1 :FREQ
?
:SENS
:MAX