7SENSe Subsystem
187 U2000 Series Programming Guide
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut]
[:MAGNitude] <numeric_value>
This command is used to enter a channel offset value for the power sensor
setup, for example cable loss. The power sensor then corrects every
measurement by this factor to compensate for the gain/loss.
Entering a value for GAIN2 using this command automatically turns the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command to ON.
Syntax
SENS
1
CORR
SENS2
:
:INP
:MAGN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:GAIN2