EasyManua.ls Logo

Agilent Technologies U2000 Series

Agilent Technologies U2000 Series
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Power Sensor Remote Operation
36 U2000 Series Programming Guide
Status Reporting
Status reporting is used to monitor the power sensor to determine when
events have occurred. Status reporting is accomplished by configuring and
reading status registers.
The power sensor has the following main registers:
•Status Register
Standard Event Register
Operation Status Register
•Questionable Status Register
Device Status Register
There are other registers that exist “behind” the main registers, and are
described later in this chapter.
Status and Standard Event registers are read using the IEEE- 488.2 common
commands.
Operation and Questionable Status registers are read using the SCPI STATus
command subsystem.
The General Status Register Model
The generalized status register model shown in Figure 1- 6 is the building
block of the SCPI status system. This model consists of a condition register, a
transition filter, an event register and an enable register. A set of these
registers is called a status group.

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Related product manuals