Power Sensor Remote Operation 1
U2000 Series Programming Guide 51
Using Device Clear to Halt Measurements
Device clear is an IEEE- 488 low- level bus message which can be used to halt
measurements in progress. Different programming languages and IEEE- 488
interface cards provide access to this capability through their own unique
commands. The status registers, the error queue, and all configuration states
are left unchanged when a device clear message is received. Device clear
performs the following actions.
All measurements in progress are aborted.
The power sensor returns to the trigger “idle state”.
The power sensor’s input and output buffers are cleared.
The power sensor is prepared to accept a new command string.
NOTE
For interfaces the that do not support a low-level device clear, use the ABORt command.