9 STATus Subsystem
234 U2000 Series Programming Guide
STATus:OPERation:CALibrating[:SUMMary]
The operation status calibrating summary register set contains information on
the calibrating status of the power sensor.
The following bits in these registers are used by the power sensor:
These bits are set at the beginning of zeroing (CALibration:ZERO:AUTO
ONCE) and at the beginning of calibration (CALibration:AUTO ONCE). Also for
the compound command/query CALibration[:ALL]?, this bit is set at the
beginning of the calibration sequence.
These bits are cleared at the end of zeroing or calibration.
Syntax
Bit Number Decimal
Weight
Definition
0- Not used
1 2 CALibrating Status
24Not used
3-15 - Not used
STAT
:OPER
:CAL
:SUMM