1 Power Sensor Remote Operation
26 U2000 Series Programming Guide
Filter Length
You specify the filter length using the following command:
[SENSe[1]]:AVERage:COUNt <numeric_value>
The range of values for the filter length is 1 to 2048. Specifying this command
disables automatic filter length selection. Increasing the value of the filter
length reduces measurement noise. However, the time to take the
measurement is increased.
Range
With an U2000 Series USB Power Sensor the range can be set either
automatically or manually. Use autoranging when you are not sure of the
power level you will be measuring.
Setting the Range
To set the range manually use the following command:
[SENSe[1]]:POWer:AC:RANGe <numeric_value>
If the <numeric_value> is set to:
• 0, the sensor’s lower range is selected.
• 1, the sensor’s upper range is selected.
For detailed on the range limits of U2000 Series USB power sensors, refer to
the U2000 Series USB Power Sensors Operating and Service Guide.
For further information on this command refer to page 198.
To enable autoranging use the following command:
[SENSe[1]]:POWer:AC:RANGe:AUTO ON
Use autoranging when you are not sure of the power level you will be
measuring.

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U2000 Series and is the answer not in the manual?

Agilent Technologies U2000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelU2000 Series
CategoryAccessories
LanguageEnglish

Related product manuals