13 IEEE 488.2 Command Reference
298 U2000 Series Programming Guide
*ESE <NRf>
The *ESE (Event Status Enable) <NRf> command sets the Standard Event
Status Enable Register. This register contains a mask value for the bits to be
enabled in the Standard Event Status Register. A 1 in the Enable Register
enables the corresponding bit in the Status Register, a 0 disables the bit. The
parameter value, when rounded to an integer and expressed in base 2,
represents the bit values of the Standard Event Status Enable Register.
Table 13- 1 shows the contents of this register.
Tab le 13-1 *ESE Mapping
Syntax
Bit Weight Meaning
0 1 Operation Complete
1 2 Request Control (not used)
2 4 Query Error
3 8 Device Dependent Error
4 16 Execution Error
5 32 Command Error
6 64 Not used
7128 Power On
*ESE
NRf
?
Space

Table of Contents