1 Power Sensor Remote Operation
34 U2000 Series Programming Guide
Fast Mode
In the highest speed setting, the limiting factor tends to be the speed of the
controller being used to retrieve results from the power sensor, and to a
certain extent, the volume of remote traffic. The latter can be reduced using
the FORMat REAL command to return results in binary format. The former is a
combination of two factors:
• the hardware platform being used
• the programming environment being used