4 CALibration Subsystem
110 U2000 Series Programming Guide
CALibration[1][:ALL]?
This query causes the power sensor to perform a zeroing sequence. The
zeroing sequence consists of:
1 Zeroing the power sensor (CALibration:ZERO:AUTO ONCE or
CALibration:AUTO ONCE)
When the zeroing sequence is completed, 0 or 1 is entered into the output
buffer to indicate if the sequence was successful. If the result is:
0, the zeroing has passed
1, the zeroing has failed
Syntax
Query Example
NOTE
This query is identical to CALibration[1][:ALL], however, unlike the command, it
provides a response to indicate whether the calibration has been successful or not.
CAL1:ALL? This command causes the power sensor to
perform a zeroing sequence and return a
result.
:ALL
?
CAL
1