7SENSe Subsystem
190 U2000 Series Programming Guide
SENSe[1]:DETector:FUNCtion <character_data>
This command sets the measurement mode for U2000 Series USB power
sensors.
Syntax
Parameters
1
When measurement mode is set to average:
If TRIG:SOUR is set to INT1, or EXT, it is set automatically to IMM.
INIT:CONT is set automatically to ON.
SENS:AVER2:STAT is set automatically to OFF.
CALC:FEED is set automatically to “POW:AVG” for all CALC blocks using the current
measurement in their CALC:MATH:EXPR.
Item Description/Default Range of Values
character_data Defines the measurement mode:
AVERage: sets the sensor to average
only mode.
AVERage
1
DET
:FUNC
SENS
1
:
Space
?
character_data