TRIGger Subsystem 11
U2000 Series Programming Guide 277
TRIGger[1][:IMMediate]
This command causes a trigger to occur immediately, provided the sensor is in
the wait for trigger state. When this command is executed, the measurement
result is stored in the power sensor’s memory. Use FETCh? to place the
measurement result in the output buffer.
Syntax
Example
Error Messages
If the power sensor is not in the wait for trigger state, then
TRIGger:IMMediate causes error –211, “Trigger ignored”.
NOTE
This command performs the same function as INITiate[1]:[IMMediate].
TRIG This command causes a sensor trigger to
occur immediately.
TRIG