3CALCulate Subsystem
96 U2000 Series Programming Guide
CALCulate[1]:LIMit:LOWer[:DATA] <numeric_value>
This command enters a value for the lower test limit for the current
window/measurement used in the CALCulate[1]:LIMit:FAIL? test. The
units used are dependent on the current setting of
UNIT:POWer. When the measured value is less than the value specified in
CALCulate[1]:LIMit:LOWer[:DATA], CALCulate[1]:LIMit:FAIL? reports
a fail. When the measured value is greater than or equal to the limit, a fail is
not reported.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value A numeric value for the lower test limit:
• DEF: the default is –90.00 dBm or
–90 db
• MIN: –150 dBm or –180 dB
• MAX: +230 dBm or +200 dB
–150 to +230 dBm or
–180 to +200 dB
DEF
MIN
MAX
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:LIM
:LOW
:DATA
CALC
1