EasyManuals Logo

Agilent Technologies U2000 Series Guide

Agilent Technologies U2000 Series
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #287 background imageLoading...
Page #287 background image
TRIGger Subsystem 11
U2000 Series Programming Guide 267
INITiate[1][:IMMediate]
This command sets the power sensor in the wait for trigger state. When a
trigger is received, the measurement is taken and the result placed in the
power sensor memory. If TRIGger:SOURce is set to IMMediate the
measurement begins as soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer.
Refer to “FETCh[1]?” on page 74 for further details.
Syntax
Example
Error Messages
If the power sensor is not in the idle state or INITiate:CONTinuous is ON,
error –213, “INIT ignored” occurs.
NOTE
This command performs the same function as
INITiate:[IMMediate]:SEQuence[1].
INIT1:IMM This command places the sensor in the
wait for trigger state.
INIT
:IMM
1

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U2000 Series and is the answer not in the manual?

Agilent Technologies U2000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelU2000 Series
CategoryAccessories
LanguageEnglish

Related product manuals