7SENSe Subsystem
171 U2000 Series Programming Guide
SENSe[1]:AVERage[:STATe] <boolean>
This command is used to enable and disable averaging.
Syntax
Example
Reset Condition
On reset, averaging is OFF.
Query
SENSe[1]:AVERage[:STATe]?
The query enters a 1 or 0 into the output buffer indicating the status of
averaging.
• 1 is returned when averaging is enabled
• 0 is returned when averaging is disabled
AVER 1 This command enables averaging on
power sensor.
SENS
1
:STAT
AVER 0|OFF
1|ON
Space
?
: