4 CALibration Subsystem
108 U2000 Series Programming Guide
CALibration Subsystem
The CALibration command subsystem is used to zero the power sensor.
Zeroing of the power sensor is recommended:
•When a 5
o
C change in temperature occurs
When you change the power sensor
Every 24 hours
Prior to measuring low level signals. For example, 10 dB above the lowest
specified power for your sensor.
The following CALibration commands are overlapped commands:
CAL:ALL
CAL:AUTO
CAL:ZERO:AUTO
An overlapped command allows the instrument to continue parsing and
executing subsequent commands while it is still executing.
Keyword Parameter Form Notes Page
CALibration
[:ALL] [event; no query] page 109
[:ALL]? [event; query] page 110
:AUTO <boolean>|ONCE page 112
:ZERO
:AUTO <boolean>|ONCE page 114
:TYPE <EXTernal|INTernal> page 116