TRIGger Subsystem 11
U2000 Series Programming Guide 275
TRIGger[1]:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling- time
delay before a measurement is made.
When this command is set to:
• ON, the power sensor inserts a settling- time delay before taking the
requested measurement. This settling time allows the internal digital filter
to be updated with new values to produce valid, accurate measurement
results. The trigger with delay command allows settling time for the
internal amplifiers and filters. It does not allow time for power sensor
delay.
In cases of large power changes, the delay may not be sufficient for
complete settling. Accurate readings can be assured by taking two
successive measurements for comparison.
• OFF, the power sensor makes the measurement immediately a trigger is
received.
Syntax
TRIG
:AUTO
:DEL
0|OFF
1|ON
Space
?