TRIGger Subsystem 11
U2000 Series Programming Guide 273
INITiate[1][:IMMediate]:SEQuence[1]
This command sets the power sensor in the wait for trigger state. When a
trigger is received, the measurement is taken and the result placed in the
power sensor memory. If TRIGger:SOURce is set to IMMediate the
measurement begins as soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer.
Refer to “FETCh[1]?” on page 74 for further information.
Syntax
Example
Error Messages
If the power sensor is not in the “idle” state or INITiate:CONTinuous is ON,
error –213, “INIT ignored” occurs.
NOTE
This command performs the same function as INITiate[1]:[IMMediate].
INIT:IMM:SEQ1 This command places the sensor in
the wait for trigger state.
INIT :IMM
:SEQ
1